Latest generation of Herzog Flash Point with proven reliability, superior design for Pensky-Martens automated closed cup testing
ASTM D 93 A&B; ISO 2719 A&B; EN ISO 2719 A & B; DIN EN ISO 2719 A&B; NF EN ISO 2719 A&B; DIN 51758; IP 34 A&B; JIS K 2265
Designed to save time, money and bench space, HFP 339 Pensky Martens analyzer automatically determines flash point in strict compliance with the appropriate test method. In addition to standard test methods, the HFP 339 may be programmed with user-defined test protocols and test for samples with an unknown flash point. An alphanumeric solvent-proof keypad, large easy-to-read VF-Display and intuitive menu flow provide an exceptionally user-friendly interface.
To increase test productivity on products having an elevated flash point, the HFP 339 begins with an accelerated heating rate and automatically switches to the standard heating rate at the appropriate temperature to assure accurate results. Results are automatically corrected for barometric pressure and displayed on an easy to read HFP VF-Display. The last 9 results are memorized in the unit and can automatically output to an optional printer or computer system (LIM).
· Accurate determination of flash point
· Economic unit with low maintenance cost
· Simple, user-friendly operation
· Electric or gas ignition
· Safety and fire detection systems
· Proven, reliable operation

Makes Pensky Martens, TAG, and Abel
flash point testing easier than ever
PENSKY MARTENS
ASTM D 93 A&B, D 6751; ISO 2719 A&B; DIN 51758
TAG
ASTM D 56
ABEL
IP 170
· Run different flash point tests with a single instrument
· Stores up to 20 permanent and user-defined test programs
· Tests for samples with unknown flash point
· Automatically corrects results to standard barometric pressure
· Easy-to-use touch screen testing, programming, diagnostics and calibrations; integral results database stores up to 300 measurements
· Built-in fire detection system with audible alarm & external relay
· Thermal detection eliminates interference from water- or silicone-containing samples
HFP 360 Series Flash Point Analyzers are designed to save you time, money and bench space. Special interchangeable test modules allow you to switch from one test method to another quickly and easily. Moreover, the HFP automatically recognizes which test module is installed so you can begin standard method testing without additional system programming or configuration.
In addition to performing standard Pensky Martens, Tag and Abel testing, this versatile analyzer is easily programmed with your own customized test protocols. Test selection and other operational commands are entered via the instrument's easy-to-use touch screen display. The system even includes online help to guide you through operational procedures.
Complete, thorough documentation
Test results are displayed on the HFP's graphics readout and can also be output automatically or on demand to an optional companion printer or computer system. A typical report includes Sample ID, Date and Time, Measurement Program, User Name, and Test Results.
To further enhance the instrument's documentation capabilities, the results from up to 300 flash point tests can be stored in its integral database and recalled for statistical analysis.

Unparalleled versatility with absolute simplicity, precision, and safety for Automated Cleveland open cup testing
ASTM D 92; ISO 2592; EN ISO 2592; DIN EN ISO 2592; NF EN ISO 2592; IP 36; JIS K 2265
The Cleveland Flash Point HFP 370 determines flash and fire points of volatile materials in accordance to ASTM D 92 or equivalent methods. Automated operation and leading flash point technology, greatly increase laboratory productivity while improving repeatability and reproducibility.
An ultra-flexible design offers unparalleled versatility, enabling you to configure the instrument best suited to your application. Traceable calibrations with lock-out control, and quality verification fulfill ISO 9000 quality system requirements. The system corrects results for barometric pressure before storing the data to permanent memory. Test progress displays in real time on the easily viewable screen.
For enhanced automation, add HFP 371 Six-Position Sample Changer at any time, program new sample positions even as another test is in progress. Each of the six test cups is equipped with its own heating plate.
· User-friendly keypad programming
· Automatic test-flame lighter
· 550 results memory storage
· Statistical results analysis (AVE, MIN, MAX, SD)
· Alert messages for out-of-spec results
· Multiple safety features
· Special equipment for efficient asphalt testing
· Optional six-position sample changer

Accurate, economical, flexible flash point testing
• PENSKY MARTENS : ASTM D 93 A&B, D 6751; ISO 2719 A&B; DIN 51758
• TAG : ASTM D 56
• ABEL : IP 170
• CLEVELAND OPEN CUP (COC) : D 92, IP 36, ISO 2592, DIN EN 22592
Designed to provide years of accurate and reliable flash point results, Herzog's top-quality HFP 380 Series Closed-Cup Flash Point Testers offer a variety of factory configurations... versatility that helps you find the unit that's exactly right for your budget and testing needs. Add a field-installed flash alert with digital temperature display to further improve your testing accuracy, convenience, and productivity.
All units feature a rugged electric heater with ultra-precise manual control, mercury thermometer, and visual flash point detection, In addition, units are available with or without a bath cooling fan (Pensky-Martens) or cooling coil (TAG and Abel) and feature an electric stirrer that automatically stops during test flame/ignitor aplication.